Hemodynamics, fluids & transfusion

A novel wearable bioimpedance sensor for continuous monitoring of fluid balance: a study on isotonic hypovolemia in healthy adults

J Clin Monit ComputPublished Experimental study

Authors' conclusion

This study demonstrates that the wearable bioimpedance device tested is very sensitive to furosemide-induced changes in fluid volume in healthy volunteers in a controlled environment.

Study context

Population
Healthy adults with furosemide-induced isotonic hypovolemia20 women · 7 men
Participants
27 adults
Outcome
Change in extracellular resistance (RE) measured by wearable bioimpedance sensor
Follow-up
Two hours after furosemide (t2)
Methodological detailssingle-group, controlled environment
  • Single-group
  • Controlled environment

Main finding

As reported in the abstract

Change in RE · t0 to t2

Resistance (RE) increased significantly from t0 to t2 (mean ± SD, %)

13.6 ± 2.9%

Further findings

Body weight reduction during hypovolemia

Body weight reduction (kg)
1.4 ± 0.2 kg
Body weight reduction (%)
1.7 ± 0.4%

Total urine elimination during fluid loss

Total urine elimination
1277 ± 190 mL

Original publication

A novel wearable bioimpedance sensor for continuous monitoring of fluid balance: a study on isotonic hypovolemia in healthy adults. J Clin Monit Comput. Published 4 December 2024. doi:10.1007/s10877-024-01245-z