Hemodynamics, fluids & transfusion
A novel wearable bioimpedance sensor for continuous monitoring of fluid balance: a study on isotonic hypovolemia in healthy adults
Authors' conclusion
This study demonstrates that the wearable bioimpedance device tested is very sensitive to furosemide-induced changes in fluid volume in healthy volunteers in a controlled environment.
Study context
- Population
- Healthy adults with furosemide-induced isotonic hypovolemia20 women · 7 men
- Participants
- 27 adults
- Outcome
- Change in extracellular resistance (RE) measured by wearable bioimpedance sensor
- Follow-up
- Two hours after furosemide (t2)
Methodological detailssingle-group, controlled environment
- Single-group
- Controlled environment
Main finding
As reported in the abstract
Change in RE · t0 to t2
Resistance (RE) increased significantly from t0 to t2 (mean ± SD, %)
13.6 ± 2.9%
Further findings
Body weight reduction during hypovolemia
- Body weight reduction (kg)
- 1.4 ± 0.2 kg
- Body weight reduction (%)
- 1.7 ± 0.4%
Total urine elimination during fluid loss
- Total urine elimination
- 1277 ± 190 mL
Original publication
A novel wearable bioimpedance sensor for continuous monitoring of fluid balance: a study on isotonic hypovolemia in healthy adults. J Clin Monit Comput. Published 4 December 2024. doi:10. 1007/ s10877-024-01245-z
Composed by AI from the published abstract and checked against it. How Briefs are made →