Hemodynamics, fluids & transfusion
A wearable cuffless blood pressure monitoring system: real-time measurement using force sensing and waveform analysis
Authors' conclusion
This study represents the first clinical validation of a tactile force-sensing cuffless system against intra-arterial reference measurements, demonstrating its feasibility for real-time, non-invasive blood pressure monitoring during anesthesia.
Study context
- Population
- Patients in the operating room
- Participants
- 46 patients
- Comparison
- Cuffless device vs Intra-arterial catheter
- Outcome
- Error between paired measurements (systolic and diastolic BP) vs intra-arterial catheter
Methodological detailsclinical validation, intra-arterial reference, Data analysis also considered the IEEE 1708 standard
- Clinical validation
- Intra-arterial reference
- Data analysis also considered the IEEE 1708 standard
Performance
As reported in the abstract
Error between paired measurements
Error (systolic)
- 1.44 ± 6.64 mmHg
Cuffless device
Error (diastolic)
3.35 ± 6.91 mmHg
Cuffless device
Further findings
Pearson correlation with intra-arterial catheter
- Systolic r
- 0.90
- Diastolic r
- 0.78
- p < 0.01 for both
Original publication
A wearable cuffless blood pressure monitoring system: real-time measurement using force sensing and waveform analysis. J Clin Monit Comput. Published 11 June 2026. doi:10. 1007/ s10877-026-01454-8
Composed by AI from the published abstract and checked against it. How Briefs are made →