A wearable cuffless blood pressure monitoring system: real-time measurement using force sensing and waveform analysis

J Clin Monit ComputPublished Method comparison (validation) study

Authors' conclusion

This study represents the first clinical validation of a tactile force-sensing cuffless system against intra-arterial reference measurements, demonstrating its feasibility for real-time, non-invasive blood pressure monitoring during anesthesia.

Study context

Population
Patients in the operating room
Participants
46 patients
Comparison
Cuffless device vs Intra-arterial catheter
Outcome
Error between paired measurements (systolic and diastolic BP) vs intra-arterial catheter
Methodological detailsclinical validation, intra-arterial reference, Data analysis also considered the IEEE 1708 standard
  • Clinical validation
  • Intra-arterial reference
  • Data analysis also considered the IEEE 1708 standard

Performance

As reported in the abstract

Error between paired measurements

Error (systolic)

- 1.44 ± 6.64 mmHg

Cuffless device

Error (diastolic)

3.35 ± 6.91 mmHg

Cuffless device

Further findings

Pearson correlation with intra-arterial catheter

Systolic r
0.90
Diastolic r
0.78
  • p < 0.01 for both

Original publication

A wearable cuffless blood pressure monitoring system: real-time measurement using force sensing and waveform analysis. J Clin Monit Comput. Published 11 June 2026. doi:10.1007/s10877-026-01454-8