Paediatric anaesthesia & critical care

Accuracy of Bispectral Index™ monitoring in an alternative nasal position in pediatric patients under general anesthesia: a prospective observational study

Can J AnaesthPublished Prospective observational study

Main finding · Agreement between nasal and frontal BIS valuesResult — go to the resultMean difference 1.0

Authors' conclusion

Bispectral Index values obtained from the nasal region in pediatric patients agreed within ±10 units of the values from conventional frontal placement.

Study context

Population
Children aged 4–12 yr undergoing GA with sevoflurane5,049 paired measurements analysed
Participants
40 children
Exposure
Nasal vs frontal BIS sensor placement
Outcome
Agreement between nasal and frontal BIS values
Methodological detailsmethod-comparison (agreement) study
  • Method-comparison (agreement) study

Main finding

As reported in the abstract

Mean difference

1.0

Concordance correlation coefficient
0.88
Marginal incidence of BIS values exceeding ±10-unit threshold
3.9% (95% CI, 2.1% to 5.8%)

Observational data: associations, not effects.

Further findings

BIS values exceeding ±10-unit threshold, emergence (subgroup)

Marginal incidence
6.6% (95% CI, 1.3% to 11.8%)

BIS values exceeding ±10-unit threshold, high-EMG (subgroup)

Marginal incidence
8% (95% CI, 1% to 15%)

BIS values exceeding ±10-unit threshold, age groups (subgroup)

Similar across age groups

Power spectral density

Spectral characteristics comparable, with lower total power at the nasal position

Original publication

Accuracy of Bispectral Index™ monitoring in an alternative nasal position in pediatric patients under general anesthesia: a prospective observational study. Can J Anaesth. Published 19 August 2026. doi:10.1007/s12630-026-03181-9

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.