Accuracy of continuous glucose monitoring systems in intensive care unit patients: a scoping review

Intensive Care MedPublished Scoping review

Authors' conclusion

In this scoping review of CGM accuracy in the ICU, we found great diversity in accuracy reporting. Accuracy varied depending on CGM and comparator, and may be better for intravascular CGM and potentially lower during hypoglycemia.

Study context

Population
Intensive care unit (ICU) patientsstudies included
Participants
96 patients
Comparison
Subcutaneous Continuous Glucose Monitoring (CGM) vs Factory calibrated CGM vs Intravenous CGM vs Intraarterial CGM
Outcome
CGM accuracy (mean absolute relative difference, mean absolute relative difference (MARD))
Methodological detailsPRISMA-ScR, PubMed and EMBASE searched; 2133 studies identified
  • PRISMA-ScR
  • PubMed and EMBASE searched; 2133 studies identified

Main finding

As reported in the abstract

MARD

Subcutaneous CGM

6.6 to 30.5%

range

Factory calibrated CGM

9.7 to 20.6%

range

Intravenous CGM

5-14.2%

Intraarterial CGM

6.4-13%

Further findings

Study characteristics

Observational studies
91.7%
Adult patients
74%
Mixed ICUs
47.9%
Subcutaneous CGM systems assessed
80%
Arterial blood samples as reference test
40.6%
Prespecified reference test protocol mentioned
56.3%

Original publication

Accuracy of continuous glucose monitoring systems in intensive care unit patients: a scoping review. Intensive Care Med. Published 17 October 2024. doi:10.1007/s00134-024-07663-6

Composed by AI from the published abstract and checked against it. How Briefs are made →