Cardiothoracic & vascular anesthesia

Benign Intravalvular Strut Leak After Placement of the Inspiris Resilia Valve

J Cardiothorac Vasc AnesthPublished Retrospective chart review

Authors' conclusion

This pattern of intravalvular leak is unique to the Inspiris Resilia valve and is commonly found in the intraoperative period.

Study context

Population
Adults undergoing aortic valve replacement with the Inspiris Resilia valve at a single tertiary care academic centre (paediatric patients and aortic valve conduits excluded)charts and echo images reviewed
Participants
81 adults
Outcome
Incidence and severity of intravalvular (strut) leak on intraoperative TEE
Follow-up
At the time of implantation
Methodological detailssingle-centre, transesophageal echocardiography (TEE) images reviewed independently by 2 echocardiographers
  • Single-centre
  • TEE images reviewed independently by 2 echocardiographers

Main finding

As reported in the abstract

Intravalvular strut leak at implantation (intraoperative TEE)

TEEs with strut leak, of 81 reviewed

56

69.1%

Trace regurgitation

30

Mild regurgitation

21

Moderate regurgitation

5

Observational data: associations, not effects.

Further findings

Persistent leak on follow-up transthoracic echocardiography · outpatient follow-up

Follow-up reports available
50
No persistent leaks

Safety

Return to CPB owing to persistent intravalvular leak

Cases necessitating return to CPB
1

Original publication

Benign Intravalvular Strut Leak After Placement of the Inspiris Resilia Valve. J Cardiothorac Vasc Anesth. Published 7 July 2024. doi:10.1053/j.jvca.2024.07.004

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.