Respiratory failure & ventilation
Clinical validation of a novel accelerometer-based respiratory rate monitor
Authors' conclusion
The wireless RespX accelerometer-based sensor provides a novel and accurate method for assessing the respiratory rate in awake hospitalised patients compared with capnography.
Study context
- Population
- Hospitalised adults (> 18 years) expected to remain undisturbed for one hourcompleted the study
- Participants
- 29 adults
- Exposure
- RespX accelerometer-based sensor vs capnography
- Outcome
- RR measured by RespX within ± 3 breaths per minute of capnography
Methodological detailssingle-centre
- Single-centre
Main finding
As reported in the abstract
Proportion of RespX measurements within ± 3 breaths/min of capnography
Proportion within ± 3 breaths per minute
77%
95% CI 74.8–79.0%
raw data
Proportion within ± 3 breaths per minute
82.8%
95% CI 80.7–84.7%
after removing outliers
Observational data: associations, not effects.
Further findings
| Outcome | Mean difference | P value |
|---|---|---|
| Mean difference in RR, RespX vs capnography (linear mixed-effects models)RespX vs Capnography · raw data · supporting their equivalence | −0.40 | P = 0.43 |
| Mean difference in RR, RespX vs capnography (linear mixed-effects models)RespX vs Capnography · filtered data · supporting their equivalence | 0.39 | P = 0.15 |
Measurement variability
- Lower measurement variability with RespX vs capnography
Safety
- No adverse events observed
Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.
Original publication
Clinical validation of a novel accelerometer-based respiratory rate monitor. J Clin Monit Comput. Published 7 February 2026. doi:10. 1007/ s10877-026-01419-x