Respiratory failure & ventilation

Clinical validation of a novel accelerometer-based respiratory rate monitor

J Clin Monit ComputPublished Prospective observational study

Authors' conclusion

The wireless RespX accelerometer-based sensor provides a novel and accurate method for assessing the respiratory rate in awake hospitalised patients compared with capnography.

Study context

Population
Hospitalised adults (> 18 years) expected to remain undisturbed for one hourcompleted the study
Participants
29 adults
Exposure
RespX accelerometer-based sensor vs capnography
Outcome
RR measured by RespX within ± 3 breaths per minute of capnography
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Proportion of RespX measurements within ± 3 breaths/min of capnography

Proportion within ± 3 breaths per minute

77%

95% CI 74.8–79.0%

raw data

Proportion within ± 3 breaths per minute

82.8%

95% CI 80.7–84.7%

after removing outliers

Observational data: associations, not effects.

Further findings

OutcomeMean differenceP value
Mean difference in RR, RespX vs capnography (linear mixed-effects models)RespX vs Capnography · raw data · supporting their equivalence−0.40P = 0.43
Mean difference in RR, RespX vs capnography (linear mixed-effects models)RespX vs Capnography · filtered data · supporting their equivalence0.39P = 0.15

Measurement variability

Lower measurement variability with RespX vs capnography

Safety

No adverse events observed

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Clinical validation of a novel accelerometer-based respiratory rate monitor. J Clin Monit Comput. Published 7 February 2026. doi:10.1007/s10877-026-01419-x

Composed by AI from the published abstract and checked against it. How Briefs are made →