Measurement accuracy of a microwave doppler sensor beneath the mattress as a continuous respiratory rate monitor: a method comparison study

J Clin Monit ComputPublished Prospective observational method comparison study

Authors' conclusion

The microwave Doppler sensor had a small bias but relatively low precision, similar to other devices. In CEG analyses, the risk of each monitor leading to inadequate clinical decision-making was low.

Study context

Population
ICU patients2,307 data points analysed
Participants
18 patients
Exposure
Microwave Doppler sensor beneath the mattress, thoracic impedance pneumography and piezoelectric sensor vs capnography (standard reference) for respiratory rate
Outcome
Measurement accuracy of respiratory rate vs capnography (bias and 95% limits of agreement)
Methodological detailsBland-Altman analysis for repeated measures, Clarke error grid analysis
  • Bland-Altman analysis for repeated measures
  • Clarke error grid analysis

Main finding

As reported in the abstract

Bias (95% limits of agreement) vs capnography

Observational data: associations, not effects.

Further findings

Correct clinical decisions (Clarke error grid analysis) — Microwave Doppler sensor
98.1%

Correct clinical decisions (Clarke error grid analysis)

Similar to the performance of the other devices

Original publication

Measurement accuracy of a microwave doppler sensor beneath the mattress as a continuous respiratory rate monitor: a method comparison study. J Clin Monit Comput. Published 4 October 2023. doi:10.1007/s10877-023-01081-7

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.