Anesthetic drugs & techniques

Comparison of automated and manual control methods in minimal flow anesthesia

J Clin Monit ComputPublished Randomised controlled trial

Authors' conclusion

The ACA mode was more advantageous than the MCA mode, reaching target AA concentrations faster and requiring fewer adjustments to achieve a constant depth of anesthesia.

Study context

Population
Patients undergoing gynecological surgery
Participants
76 patients
Comparison
Automatically controlled anesthesia (ACA) vs MCA

Primary result

As reported in the abstract

Depth of anesthesia, anesthetic agent (AA) consumption, time to target, awakening times, ventilator adjustments
  • No statistically significant difference in depth of anesthesia
  • No primary endpoint named in the abstract

Further findings

Time to achieve MAC target of 1

OutcomeACAMCAP value
Anesthetic agent consumptionNo statistically significant difference19.1 ± 4.9 ml17.2 ± 4.5P = 0.076
Number of vaporizer adjustments15217

Time to achieve minimum alveolar concentration (MAC) target of 1

ACA mode achieved the MAC target significantly faster

Original publication

Comparison of automated and manual control methods in minimal flow anesthesia. J Clin Monit Comput. Published 25 April 2024. doi:10.1007/s10877-024-01163-0

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.