Sepsis & infection

Effectiveness of Continuous Automatic Endotracheal Cuff Pressure Monitoring in Preventing Ventilator-Associated Pneumonia in the Pediatric Intensive Care Unit: A Randomized Controlled Trial

ChestPublished open-label, randomized controlled superiority trial

Primary result · Proportion of patients developing VAPResult — go to the resultRisk difference −10.6 percentage points 95% CI −20.3 to −0.8

Authors' conclusion

CACPM significantly reduces the proportion of critically ill pediatric patients developing VAP compared with IMCPM. This preventive effect appears directly linked to improved target pressure compliance, suggesting that automated regulation could be a valuable addition to pediatric VAP prevention strategies.

Study context

Population
Critically ill pediatric patients requiring mechanical ventilation for ≥48 hours
Participants
241 randomized patients
Comparison
Continuous automatic cuff pressure monitoring (CACPM) compared to intermittent manual cuff pressure monitoring (IMCPM)
Primary endpoint
Proportion of patients developing VAP
Follow-up
Within 28 days following intubation
Methodological details1:1 randomization, intention to treat analysis
  • 1:1 randomization
  • Intention to treat analysis

Primary result

As reported in the abstract

Risk difference

−10.6 percentage points

95% CI −20.3 to −0.8

P = 0.034

By group

CACPM vs IMCPM
risk difference −10.6 percentage points (95% CI −20.3 to −0.8); P = 0.034

Secondary outcomes

Proportion of patients developing VAP, per-protocol analysis · PP analysis

P = 0.044

OutcomeCACPMIMCPMEstimateP value
VAP incidence densityOrder of groups for the IRR not printed · ITT analysis15.7 per 1,000 ventilator-days27.1 per 1,000 ventilator-daysIRR 1.73 (95% CI 0.94–3.18)P = 0.080
Time to VAP diagnosisOrder of groups not printed · multivariable Cox proportional hazard regressionHR for VAP, unadjusted 1.80 (95% CI 0.976–3.321)P = 0.060
Time to VAP diagnosisHigher VAP hazard with IMCPM · multivariable Cox proportional hazard regressionHR for VAP, adjusted 1.90 (95% CI 1.024–3.510)P = 0.042
OutcomeP value
Target-pressure complianceHigher with CACPMP < 0.001

Original publication

Effectiveness of Continuous Automatic Endotracheal Cuff Pressure Monitoring in Preventing Ventilator-Associated Pneumonia in the Pediatric Intensive Care Unit: A Randomized Controlled Trial. Chest. Published 10 October 2026. doi:10.1016/j.chest.2026.10.003

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.