Evaluation of point-of-care haemoglobin measurement accuracy in surgery (PREMISE) and implications for transfusion practice: a prospective cohort study

Br J AnaesthPublished Prospective diagnostic cohort study

Authors' conclusion

No point-of-care Hgb device demonstrated limits of agreement that were smaller than the agreement difference of 4 g L-1.

Study context

Population
Consecutive patients undergoing noncardiac surgery and requiring point-of-care haemoglobin (Hgb) measurements1,735 intraoperative samples
Participants
1,139 patients
Comparison
Point-of-care Hgb devices vs laboratory Hgb
Outcome
Individual pairwise comparisons between point-of-care and laboratory Hgb values; agreement threshold within 4 g L-1 (limits of agreement)
Methodological detailsmethod comparison, two academic hospitals
  • Method comparison
  • Two academic hospitals

Performance

As reported in the abstract

Limits of agreement vs laboratory Hgb, samples with Hgb <100 g L-1
Test or modelLimits of agreement
HemoCue®−9.5 to 8.0
i-STAT®−16.2 to 11.5
Rad-67®−14.7 to 40.5

Limits of agreement

HemoCue®−9.5 to 8.0
i-STAT®−16.2 to 11.5
Rad-67®−14.7 to 40.5

Further findings

OutcomeHemoCue®i-STAT®Rad-67®
Potentially clinically significant transfusion error5.8%25.3%28.2%
Hgb measurements within 10 g L-1 of laboratory valueIntraoperative blood samples98%— not stated in the abstract— not stated in the abstract

— not stated in the abstract

Original publication

Evaluation of point-of-care haemoglobin measurement accuracy in surgery (PREMISE) and implications for transfusion practice: a prospective cohort study. Br J Anaesth. Published 9 January 2025. doi:10.1016/j.bja.2024.09.033

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.