Hemodynamics, fluids & transfusion

Trending Ability and Accuracy of Minimally Invasive Pulse Wave Analysis Devices: A Systematic Review and Meta-Analysis

Anesth AnalgPublished Systematic review and meta-analysis of method-comparison studies

Authors' conclusion

Minimally invasive pulse wave analysis devices do not achieve sufficient accuracy or trending ability to be interchangeable with reference techniques.

Study context

Population
Studies of perioperative and intensive care unit (ICU) hemodynamic monitoring comparing minimally invasive pulse wave analysis devices with reference techniques50 studies
Comparison
Pulse wave analysis devices vs Reference techniques
Outcome
Accuracy/precision (bias, percentage error) and trending ability (four-quadrant concordance, polar plot agreement) of CO measurement
Methodological detailsrandom-effects models, predefined fitness appraisal, subgroup analyses, sensitivity analyses, 49 studies provided accuracy/precision data (69 lines of analysis); all 50 provided trending ability data (70 lines)
  • Random-effects models
  • Predefined fitness appraisal
  • Subgroup analyses
  • Sensitivity analyses
  • 49 studies provided accuracy/precision data (69 lines of analysis); all 50 provided trending ability data (70 lines)

Pooled result

As reported in the abstract

Accuracy and trending ability vs reference techniques (all devices and settings)

Pooled bias

0.09 L/min

95% CI −0.05 to 0.23

Pooled percentage error

48.3%

unweighted mean 49.0%, 95% CI, 45.8–52.2

Pooled four-quadrant concordance

72.6%

95% CI 67.9–77.4

Polar plot ±30° agreement

63.2%

95% CI 54.9–71.5

Further pooled results

Percentage error by clinical setting (subgroup)

Lower percentage error in ICU than in noncardiac surgery
≈42% vs ≈55%

Results by declared conflict of interest (subgroup)

More favorable results in studies with declared conflict of interest
≈44% vs 51%, P < .05

MostCare and Argos trending ability (subgroup)

Four-quadrant concordance >90% and polar agreement close to accepted thresholds; based on few studies

High-quality datasets only (sensitivity analysis)

Results very similar to the overall estimates

Original publication

Trending Ability and Accuracy of Minimally Invasive Pulse Wave Analysis Devices: A Systematic Review and Meta-Analysis. Anesth Analg. Published 7 April 2026. doi:10.1213/ane.0000000000008008

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.