Respiratory failure & ventilation

High‐flow via a tracheostomy tube and speaking valve during weaning from mechanical ventilation and tracheostomy

Acta Anaesthesiol ScandPublished Retrospective register study with benchtop experiments

Authors' conclusion

Current study clinical data and benchtop experiments indicate that HFT-T-OWV was well tolerated and appeared safe.

Study context

Population
Tracheostomized patients undergoing weaning attempts at an intensive care rehabilitation center, 2019–2022124 treated with HFT-T-OWV
Participants
128 patients
Outcome
Adverse events of HFT-T-OWV
Methodological detailssingle-centre, quality register, benchtop simulation
  • Single-centre
  • Quality register
  • Benchtop simulation

Main finding

As reported in the abstract

Adverse events related to HFT-T-OWV

HFT-T-OWV

0

events

Observational data: associations, not effects.

Further findings

Main reason for not using HFT-T-OWV

Partial upper airway obstruction using a one-way valve (OWV)

Benchtop: HFT-T-OWV maximum pressure and pressure support (25-60 L/min flow)

Maximum pressures
<4 cmH2 O
Pressure support
0-0.6 cmH2 O

Benchtop: standard OWV with 1-3 L/min supplemental oxygen, simulated airway obstruction

Pressures
between 84 and 148 cmH2 O

Safety

OutcomeValues
Adverse events related to HFT-T-OWVHFT-T-OWV 0 events

Original publication

High‐flow via a tracheostomy tube and speaking valve during weaning from mechanical ventilation and tracheostomy. Acta Anaesthesiol Scand. Published 12 July 2023. doi:10.1111/aas.14305

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.