Real-time changes in pulse pressure during a 10-second end-expiratory occlusion test reliably detect preload responsiveness

Crit CarePublished Diagnostic accuracy study

An erratum to this article was published (Crit Care, 19 Aug 2025 ↗)

Authors' conclusion

In ventilated patients with sinus rhythm, real-time changes in PP during a 10-second EEO reliably detect preload responsiveness.

Study context

Population
Mechanically ventilated patients with sinus rhythm; preload responsiveness = fluid bolus-induced ΔCI ≥ 15% or passive leg raising-induced ΔCI ≥ 10%61 (43%) preload responders
Participants
143 participants
Comparison
End-expiratory occlusion (EEO)-induced ∆PP and ∆CI at 15-, 10- and 5-second durations, with or without end-inspiratory occlusion (EIO), for detecting preload responsiveness
Outcome
Diagnostic accuracy (area under the receiver operating characteristic (AUROC)) for detecting preload responsiveness

Performance

As reported in the abstract

AUROC for detecting preload responsiveness
Test or modelAUC
∆CIEEO-15''0.935 [0.881-0.969]
∆CIEEO-10''0.910 [0.851-0.951]
∆CIEEO-5''0.541 [0.456-0.625]
∆PPEEO-15''0.913 [0.857-0.952]
∆PPEEO-10''0.912 [0.860-0.947]
∆PPEEO-5''0.834 (0.763-0.891

AUC

∆CIEEO-15''0.935 [0.881-0.969]
∆CIEEO-10''0.910 [0.851-0.951]
∆CIEEO-5''0.541 [0.456-0.625]
∆PPEEO-15''0.913 [0.857-0.952]
∆PPEEO-10''0.912 [0.860-0.947]
∆PPEEO-5''0.834 (0.763-0.891
  • ∆CIEEO-10'' vs ∆CIEEO-15'': P = 0.410 — similar
  • ∆CIEEO-5'' vs ∆CIEEO-15'': P < 0.001 — smaller
  • ∆PPEEO-15'' vs ∆CIEEO-15'': P = 0.346 — similar
  • ∆PPEEO-10'' vs ∆CIEEO-15'': P = 0.336 — similar
  • ∆PPEEO-5'' vs ∆CIEEO-15'': P = 0.005 — significantly smaller

Further findings

EEO-induced changes, responders vs non-responders

∆CIEEO-15'' responders vs non-responders (median [IQR])
5.8 [4.5-7.3]% vs. 1.1 [0.1-3.4]%; p < 0.001
∆CIEEO-10'' responders vs non-responders
3.0 [2.4-4.3]% vs. 0.6 [0.1-1.6]%; p < 0.001
∆PPEEO-5'' responders vs non-responders
5.2 [2.8-8.7]% vs. 1.2 [0.3-2.8]%; p < 0.001
∆PPEEO-10'' responders vs non-responders
7.7 [5.0-12.4]% vs. 1.8 [0.5-3.1]%; p < 0.001
∆PPEEO-15'' responders vs non-responders
8.1 [5.1-11.8]% vs. 1.5 [0.5-3.0]%; p < 0.001
∆CIEEO-5'' did not differ between responders and non-responders

Adding EIO (∆CIEEO+EIO, ∆PPEEO+EIO)

Did not enhance reliability of the test at each test duration

Original publication

Real-time changes in pulse pressure during a 10-second end-expiratory occlusion test reliably detect preload responsiveness. Crit Care. Published 13 June 2025. doi:10.1186/s13054-025-05483-8