Respiratory failure & ventilation
Regional ventilation dynamics of electrical impedance tomography validated with four-dimensional computed tomography: single-center, prospective, observational study
Authors' conclusion
Regional ventilation using electrical impedance tomography during dynamic ventilation was highly accurate and consistent with the time phase compared to four-dimensional computed tomography.
Study context
- Population
- Adults receiving passive mechanical ventilation in a general ICU, July 2021 to February 2024analysed
- Participants
- 45 adults
- Exposure
- Electrical impedance tomography vs Four-dimensional CT
- Outcome
- Regional accuracy (correlation of regional dynamic ventilation) of electrical impedance tomography vs four-dimensional CT
Methodological detailssingle-center, validation against four-dimensional CT
- Single-center
- Validation against four-dimensional CT
Main finding
As reported in the abstract
Correlation coefficients, regional dynamic ventilation
- Right lung, ventral
- 0.963
- Right lung, central
- 0.963
- Right lung, dorsal
- 0.835
- Left lung, ventral
- 0.947
- Left lung, central
- 0.927
- Left lung, dorsal
- 0.823
Observational data: associations, not effects.
Further findings
Correlation coefficients after excluding 9 patients with regional ventilation distribution < 2%
- Right lung, ventral
- 0.963
- Right lung, central
- 0.963
- Right lung, dorsal
- 0.946
- Left lung, ventral
- 0.942
- Left lung, central
- 0.924
- Left lung, dorsal
- 0.951
Original publication
Regional ventilation dynamics of electrical impedance tomography validated with four-dimensional computed tomography: single-center, prospective, observational study. Crit Care. Published 16 October 2024. doi:10. 1186/ s13054-024-05130-8