Regional ventilation dynamics of electrical impedance tomography validated with four-dimensional computed tomography: single-center, prospective, observational study

Crit CarePublished Prospective observational study

Authors' conclusion

Regional ventilation using electrical impedance tomography during dynamic ventilation was highly accurate and consistent with the time phase compared to four-dimensional computed tomography.

Study context

Population
Adults receiving passive mechanical ventilation in a general ICU, July 2021 to February 2024analysed
Participants
45 adults
Exposure
Electrical impedance tomography vs Four-dimensional CT
Outcome
Regional accuracy (correlation of regional dynamic ventilation) of electrical impedance tomography vs four-dimensional CT
Methodological detailssingle-center, validation against four-dimensional CT
  • Single-center
  • Validation against four-dimensional CT

Main finding

As reported in the abstract

Correlation coefficients, regional dynamic ventilation
Right lung, ventral
0.963
Right lung, central
0.963
Right lung, dorsal
0.835
Left lung, ventral
0.947
Left lung, central
0.927
Left lung, dorsal
0.823

Observational data: associations, not effects.

Further findings

Correlation coefficients after excluding 9 patients with regional ventilation distribution < 2%

Right lung, ventral
0.963
Right lung, central
0.963
Right lung, dorsal
0.946
Left lung, ventral
0.942
Left lung, central
0.924
Left lung, dorsal
0.951

Original publication

Regional ventilation dynamics of electrical impedance tomography validated with four-dimensional computed tomography: single-center, prospective, observational study. Crit Care. Published 16 October 2024. doi:10.1186/s13054-024-05130-8

Composed by AI from the published abstract and checked against it. How Briefs are made →