Respiratory failure & ventilation

The evaluation of a non-invasive respiratory monitor in ards patients in supine and prone position

J Clin Monit ComputPublished Prospective observational study

Authors' conclusion

The RVM is accurate in assessing tidal volume and respiratory rate in prone compared to supine position.

Study context

Population
Sedated, paralysed, mechanically ventilated (volume-controlled) ARDS patients scheduled for prone positioning
Participants
40 participants
Exposure
Respiratory volume monitor (RVM) measurements in supine vs prone position
Outcome
Accuracy of impedance-based non-invasive respiratory volume monitor (RVM) for tidal volume and respiratory rate

Main finding

As reported in the abstract

Tidal volume and respiratory rate, supine vs prone

Bias (limits of agreement)

0.12 bpm

CI −1.4 to 1.6

respiratory rate

Bias (limits of agreement)

20 mL

CI −80 to 120

tidal volume

No significant difference supine vs prone for tidal volume
P = 0.795; p = 0.302
No significant difference supine vs prone for respiratory rate
P = 0.181; p = 0.604

Observational data: associations, not effects.

Original publication

The evaluation of a non-invasive respiratory monitor in ards patients in supine and prone position. J Clin Monit Comput. Published 26 March 2024. doi:10.1007/s10877-024-01147-0