Bioimpedance as a measure of fluids in patients with septic shock. A prospective observational study

Acta Anaesthesiol ScandPublished Prospective observational cohort study

Authors' conclusion

In patients with septic shock bioelectrical impedance analysis had no systematic errors or bias, but wide limits of agreement, indicating that the devices have a large and uncorrectable random error.

Study context

Population
Adult ICU patients with septic shock
Participants
49 patients
Exposure
Agreement of bioelectrical impedance analysis measures: left vs right side, 1 h apart, and between two devices
Outcome
Validity of bioelectrical impedance analysis for fluid measures (Bland Altman 95% limits of agreement; correlation with clinical markers of fluids)
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Fluid overload 1 h apart: most narrow 95% limits of agreement (L)

−2.4-2.9

Between devices: same wide limits of agreement; e.g. total body water 95% limits of agreement (L)

−14.8 −16.7

Observational data: associations, not effects.

Original publication

Bioimpedance as a measure of fluids in patients with septic shock. A prospective observational study. Acta Anaesthesiol Scand. Published 18 January 2023. doi:10.1111/aas.14180