Authors' conclusion

The raw EEG signal obtained with needle electrodes for processed EEG monitoring is comparable to that of conventional electrodes.

Study context

Population
Patients under GA for elective surgery3,447 BIS pairs analysed
Participants
23 participants
Exposure
Needle electrodes vs conventional BIS sensor
Outcome
Agreement (Bland-Altman) and correlation of BIS values between needle electrodes and conventional sensor
Methodological detailsmethod-comparison (Bland-Altman)
  • Method-comparison (Bland-Altman)

Main finding

As reported in the abstract

BIS agreement and correlation

R

0.912

95% CI 0.9057–0.9170

strong correlation

Mean difference

1.735

95% CI - 8.983 to 12.454

SD ± 5.469; interval is 95% limits of agreement

Observational data: associations, not effects.

Original publication

Comparing signal concordance of needle electrodes and conventional bispectral index sensors in anesthesia monitoring: a prospective pilot study. J Clin Monit Comput. Published 9 December 2025. doi:10.1007/s10877-025-01389-6