Cardiothoracic & vascular anesthesia

iThermonitor: A wearable non-invasive high-precision alternative to traditional temperature monitoring in thoracoscopic surgery of lateral decubitus position

J Clin Monit ComputPublished Diagnostic accuracy study

Performance · iThermonitor readings within ± 0.5 °C of lower esophageal valuesResult — go to the resultPearson r 0.75

Authors' conclusion

The iThermonitor is a reliable and accurate non-invasive alternative to traditional invasive temperature monitoring methods in thoracoscopic surgery.

Study context

Population
Adults scheduled for thoracoscopic surgery under GA, December 1, 2023, to May 31, 2024
Participants
80 adults · 3,536 temperature pairs
Comparison
iThermonitor vs Lower esophageal temperature
Outcome
Agreement between devices (± 0.5 °C)
Methodological detailsTemperatures recorded every 3 minutes
  • Temperatures recorded every 3 minutes

Performance

As reported in the abstract

iThermonitor readings within ± 0.5 °C of lower esophageal values

Pearson r

0.75

P < 0.0001

By group

iThermonitor

93.96%

95% CI 91.94–95.85%

Further findings

Diagnostic accuracy for hypothermia (<36.0 °C)

The device demonstrated high diagnostic accuracy for hypothermia; AUC
0.876

Safety

Device-related mucosal/skin effects

  • Device-related mucosal/skin effects: Esophageal probes: bloodstaining; iThermonitor: skin redness, resolved postoperatively

Original publication

iThermonitor: A wearable non-invasive high-precision alternative to traditional temperature monitoring in thoracoscopic surgery of lateral decubitus position. J Clin Monit Comput. Published 2 September 2025. doi:10.1007/s10877-025-01348-1

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.