Anesthetic drugs & techniques

Noninvasive cardiac index estimation under general anaesthesia: comparison between the VenArt® device and transthoracic echocardiography

BJA OpenPublished Method comparison study

Authors' conclusion

The VenArt® device showed negligible bias and acceptable differences. Trending ability was favourable, with clinically acceptable agreement and high concordance in tracking haemodynamic changes.

Study context

Population
Women (ASA I–III) undergoing laparoscopic gynaecological procedures under GA273 paired values analysed
Participants
55 women
Exposure
VenArt® device vs transthoracic echocardiography for cardiac index
Outcome
Agreement in cardiac index (Bland-Altman: bias, precision, mean percentage error, limits of agreement)
Methodological detailsprospective, observational
  • Prospective
  • Observational

Main finding

As reported in the abstract

Agreement in cardiac index between VenArt® and transthoracic echocardiography (Bland-Altman analysis)
Bias (95% CI 0-0.05), L min-1 m-2
0.02
Precision
0.20
Mean percentage error
14.7% (95% CI 13.2-16.2%)
Lower limit of agreement (95% CI −0.41 to −0.33), L min-1 m-2
−0.37
Upper limit of agreement (95% CI 0.37-0.45), L min-1 m-2
0.41

Observational data: associations, not effects.

Secondary outcomes

Trending ability of cardiac index changes

Four-quadrant concordance rate
95.88%
Mean polar angle
0.75°
Polar angle SD
13.4°
Radial limits of agreement within ±30°
  • Good agreement

Original publication

Noninvasive cardiac index estimation under general anaesthesia: comparison between the VenArt® device and transthoracic echocardiography. BJA Open. Published 17 December 2025. doi:10.1016/j.bjao.2025.100516

Composed by AI from the published abstract and checked against it. How Briefs are made →