Cardiothoracic & vascular anesthesia

Point-of-care testing for tranexamic acid efficacy: a proof-of-concept study in cardiac surgical patients

Br J AnaesthPublished Mechanistic study

Authors' conclusion

The TPA-test is sensitive to low concentrations of TXA and serves as a practical monitoring tool for postoperative fibrinolytic activity in cardiac surgery patients.

Study context

Population
Healthy volunteers (in vitro) and cardiac surgery patients undergoing cardiopulmonary bypass (CPB)plus 6 volunteers in vitro
Participants
20 volunteers
Outcome
TXA inhibition of tPA-triggered fibrinolysis on the TPA-test (ClotPro®)
Methodological detailsin vitro evaluation, observational, proof-of-concept
  • In vitro evaluation
  • Observational
  • Proof-of-concept

Main finding

As reported in the abstract

In vitro: Tissue plasminogen activator (tPA)-induced hyperfibrinolysis inhibited by tranexamic acid (TXA) in a concentration-dependent manner from

≥2 mg L-1

In vitro: complete inhibition at TXA

≥10 mg L-1

Patients: antifibrinolytic effect detectable after TXA bolus at end of CPB of

0.1 g

Patients: complete inhibition of fibrinolysis with TXA

≥0.5 g

Further findings

Renal function and TPA-test parameters · 18 h after surgery

Fibrinolytic inhibition continued in patients with estimated glomerular filtration rate (eGFR) ≤30 ml min-1 1.73 m-2 (n)
4
Correlation of eGFR with TPA-test parameters (r)
0.86-0.92
  • P<0.0001

Original publication

Point-of-care testing for tranexamic acid efficacy: a proof-of-concept study in cardiac surgical patients. Br J Anaesth. Published 26 April 2024. doi:10.1016/j.bja.2024.03.023

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.