Cardiothoracic & vascular anesthesia

Accuracy of non-invasive sensors measuring core body temperature in cardiac surgery intensive care unit (ICU) patients – results from a monocentric prospective observational study

J Clin Monit ComputPublished Prospective observational study

Authors' conclusion

Core temperature was generally underestimated by the non-invasive approaches. In our study, ZHF outperformed DS. In terms of agreement, results for both sensors were outside the range that is considered clinically acceptable.

Study context

Population
Patients admitted to the ICU after cardiac surgery1,600 measurement sets
Participants
40 patients
Exposure
Double-Sensor (DS) and Zero-Heat-Flux (ZHF) forehead sensors vs pulmonary artery catheter (PAC) core temperature
Outcome
Agreement with PAC core temperature (Bland-Altman for repeated measurements)
Methodological detailsmethod comparison, monocentric, Measurements every 5 min, up to 40 data sets per patient
  • Method comparison
  • Monocentric
  • Measurements every 5 min, up to 40 data sets per patient

Main finding

As reported in the abstract

Agreement with PAC (Bland-Altman mean bias ± 95% LoA)
GroupValueLin's concordance correlation coefficient (LCCC)
DSmean bias −0.82 (95% Limits-of-Agreement (LoA) ± 1.27) °C0.5
ZHFmean bias - 0.54 (95% LoA ± 1.14) °C0.63

Value

DSmean bias −0.82 (95% LoA ± 1.27) °C
ZHFmean bias - 0.54 (95% LoA ± 1.14) °C

Lin's concordance correlation coefficient (LCCC)

DS0.5
ZHF0.63

Observational data: associations, not effects.

Further findings

OutcomeDSZHF
Detection of hyperthermia (≥ 38 °C)
Sensitivity0.120.35
Specificity0.991.0
Detection of hypothermia (< 36 °C)
Sensitivity0.951.0
Specificity0.720.85

Mean bias in hyperthermic and hypothermic patients (subgroup)

Mean bias significantly higher in hyperthermic and hypothermic patients

Original publication

Accuracy of non-invasive sensors measuring core body temperature in cardiac surgery ICU patients – results from a monocentric prospective observational study. J Clin Monit Comput. Published 12 July 2023. doi:10.1007/s10877-023-01049-7

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.