Cardiothoracic & vascular anesthesia

Intraoperative zero-heat-flux thermometry overestimates nasopharyngeal temperature by 0.39 °C: an observational study in patients undergoing congenital heart surgery

J Clin Monit ComputPublished Prospective observational study

Study context

Population
Pediatric and adult patients undergoing surgery for congenital heart disease47 patients
Exposure
Zero-heat-flux SpotOn and Tcore (Tcore), and vesical temperature, vs nasopharyngeal temperature
  • TSpotOnn = 1,073
  • Tcoren = 874
  • Tvesn = 1,102
Outcome
Accuracy and precision vs TNPH (Bland-Altman)
Methodological detailsmethod comparison (Bland-Altman)
  • Method comparison (Bland-Altman)

Main finding

As reported in the abstract

Agreement with TNPH
OutcomeTSpotOnTcoreTves
Agreement with TNPHaccuracy - 0.39 (precision - 1.12 and 0.35) °Caccuracy −0.09 (precision −0.88–0.71) °Caccuracy 0.07 (precision - 1.90 to 2.05) °C
PoD ≤ 0.5 °C71%71%60%
LCC0.94550.95100.9322

Observational data: associations, not effects.

Further findings

Lower recording limit

Not recordable non-invasively below (TSpotOn), °C
25.2
Not recordable non-invasively below (Tcore)
27.1
Lower temperatures recorded only with Tves

Original publication

Intraoperative zero-heat-flux thermometry overestimates nasopharyngeal temperature by 0.39 °C: an observational study in patients undergoing congenital heart surgery. J Clin Monit Comput. Published 10 August 2024. doi:10.1007/s10877-024-01204-8

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.