Cardiothoracic & vascular anesthesia

ClotPro Viscoelastometry Evaluation in Cardiac Surgery With Cardiopulmonary Bypass

J Cardiothorac Vasc AnesthPublished Prospective observational method-comparison study

An erratum to this article was published (J Cardiothorac Vasc Anesth, Aug 2023 ↗)

Authors' conclusion

ClotPro can be used as an alternative to ROTEM delta to evaluate coagulation function in cardiac surgery

Study context

Population
Adults undergoing cardiac surgery with cardiopulmonary bypass (CPB) at a university hospital
Participants
40 adults
Exposure
ClotPro vs ROTEM delta viscoelastometry
Outcome
Reproducibility of ClotPro and correlation of its parameters with ROTEM delta
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Reproducibility and correlation

Excellent correlation ROTEM delta vs ClotPro in most parameters

≥0.93

Moderate correlation, EXTEM clotting time vs EX-test

0.54

Observational data: associations, not effects.

Further findings

OutcomeP value
Absolute differences in test results between devicesMost measurements significantly differentP < 0.0001

Original publication

ClotPro Viscoelastometry Evaluation in Cardiac Surgery With Cardiopulmonary Bypass. J Cardiothorac Vasc Anesth. Published 25 November 2022. doi:10.1053/j.jvca.2022.11.022

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.