Cardiothoracic & vascular anesthesia

Reliability of bioreactance and arterial waveform analyses in monitoring stroke volume variation during cardiac surgery

J Clin Monit ComputPublished Prospective observational method comparison study

Authors' conclusion

Despite acceptable mean biases, the variability between devices was considerable, and trending analyses indicated only limited concordance. The studied SVV monitors, therefore, cannot be considered interchangeable in the context of cardiac surgery.

Study context

Population
Patients undergoing cardiac surgery (OPCAB or with cardiopulmonary bypass (CPB)), single university hospital2,055 + 367 paired stroke volume variation (SVV) measurements
Participants
18 vs 20
Exposure
SVV by bioreactance vs arterial waveform analysis
  • OPCABn = 18
  • CPBn = 20
Outcome
Agreement (Bland-Altman) and trending ability (four-quadrant plots, error grids, concordance) of SVV measurements between devices
Follow-up
Intraoperatively and postoperatively
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Agreement of SVV between devices

Mean bias

2.3 %pt

95% CI 2.1–2.6

Starling SV vs FloTrac · limits of agreement −14.3 to 20.5%pt (wide); OPCAB, 2055 paired measurements

Bias

1.5 %pt

95% CI 0.9–2.2

Starling SV vs LiDCOrapid · limits of agreement −38.3 to 38.4%pt (very wide); CPB, 367 paired measurements

Observational data: associations, not effects.

Original publication

Reliability of bioreactance and arterial waveform analyses in monitoring stroke volume variation during cardiac surgery. J Clin Monit Comput. Published 24 November 2025. doi:10.1007/s10877-025-01385-w

Composed by AI from the published abstract and checked against it. How Briefs are made →